Performance Characteristics
Nominal Range | 0~1000 ppm |
Maximum Overload | 2000 ppm |
Sensitivity(20℃) | 0.03 ± 0.01μA/ppm |
Response Time (T90) | ≤ 50 s |
Zero Signal(20℃) | < ± 0.2μA |
Baseline Shift (- 20℃ ~ 50℃) | < 10 ppm |
Resolution | 10 ppm |
Linearity | linear up to 1000ppm |
Bias Voltage | zero |
Product Dimensions
Environmental
Temperature Range | -20℃ ~ 50℃ |
Pressure Range | 1 atm ± 10% |
Humidity Range | 15% ~ 95% RH non-condensing |
Life time
Long time output drift | < 2% signal/month |
Recommended Storage Temp | 10℃ ~ 30℃ |
Expected Operating Life | 2 years in clean air |
Storage Life | 6 months in original packaging |
Warranty | 12 months |
Intrinsic safety Data
Maximum Current at 2000ppm H2 | < 0.2mA |
Maximum O/C Voltage | 1.3V |
Maximum S/C Current | < 1.0A |
Mechanical
Housing Material | ABS |
Weight (Nominal) | 8g |
Orientation | None |